For object and some calibration exposures we need to use the shutter times. Neither the ccd nor the hx actors do that correctly. And they do it differently. Fix.
cloomis@shell-ics:2024-06-03$ fitsheader -f -e0 -k W_CAM -k HST-STR {sps,ramps}/PFS?111737*
filename W_CAM HST-STR
----------------------- ----- ------------
sps/PFSA11173711.fits b1 19:27:22.741
sps/PFSA11173712.fits r1 19:27:22.662
sps/PFSA11173721.fits b2 19:27:46.372
sps/PFSA11173722.fits r2 19:27:46.341
sps/PFSA11173731.fits b3 19:27:40.136
sps/PFSA11173732.fits r3 19:27:40.155
sps/PFSA11173741.fits b4 19:27:22.711
sps/PFSA11173742.fits r4 19:27:22.705
ramps/PFSB11173723.fits n2 19:30:54.448
ramps/PFSB11173733.fits n3 19:30:59.277
(rubin3_ics) alefur@shell-ics:/data/logs/actors$ cat enu_sm*/2024-06-02*|grep shutterTimings=111737 2024-06-03 05:30:48.813Z cmds 20 CommandLink.py:122 > 2 3507 i shutterTimings=111737,2024-06-02T19:27:48.386778,2024-06-02T19:27:48.782632,2024-06-02T19:30:48.387398,2024-06-02T19:30:48.778487 2024-06-03 05:30:48.830Z cmds 20 CommandLink.py:122 > 2 3886 i shutterTimings=111737,2024-06-02T19:27:48.386466,2024-06-02T19:27:48.782507,2024-06-02T19:30:48.387160,2024-06-02T19:30:48.782388 2024-06-03 05:30:48.807Z cmds 20 CommandLink.py:122 > 2 3966 i shutterTimings=111737,2024-06-02T19:27:48.386243,2024-06-02T19:27:48.782049,2024-06-02T19:30:48.387169,2024-06-02T19:30:48.773816 2024-06-03 05:30:48.800Z cmds 20 CommandLink.py:122 > 2 3709 i shutterTimings=111737,2024-06-02T19:27:48.376621,2024-06-02T19:27:48.772964,2024-06-02T19:30:48.377248,2024-06-02T19:30:48.767711