update enu.yaml with new SM1 slit position

    XMLWordPrintable

    Details

    • Sprint:
      preRun26Jan2026

      Description

      Like in the November 2025 run (run 25), the SM1 slit seems to have moved again. As a short action, update home position.

      slitThroughFocus shows 

      •  defocus (shift-X): -0.1833 mm
      •  tilt-Y : 0.0471 deg
         To correct them, the home position should be updated as follow:
         
         [ -0.522 + (-0.1833), 0.097, 205.146025, 0.0, 0.059076 - 0.0471, -120 ]

      = [ -0.7053, 0.097, 205.146025, 0.0, 0.011976, -120 ]

      Note that we should inspect the SM1 slit for the future run.